THIS REPORT PREPARED UNDER DUTY IMPOSED BY A.A.C. R13-10-104(A) ARIZONA DEPARTMENT OF PUBLIC SAFETY
STANDARD QUALITY ASSURANCE PROCEDURES INTOXILYZER MODEL 5000EN
STANDARD CALIBRATION CHECK PROCEDURE
QA SPECIALIST AGENCY
DATE TIME
INTOXILYZER SERIAL # LOCATION
( ) 1. a. Ensure the dry gas tank is attached to the instrument and contains a standard alcohol concentration solution,
AC.
( ) b. Pour a standard alcohol concentration solution AC, into a clean dry simulator and assemble the simulator. Ensure that a tight seal is made. Turn on the simulator and allow temperature to reach 34° C ± 0.2° C.
( ) 2. Intoxilyzer 5000EN display reads “PUSH BUTTON ….”
( ) 3. Ensure Intoxilyzer 5000EN calibration standard is set for “G” for gas or “W” for wetbath. ( ) 4. Type “C” and press ENTER key on the keyboard.
( ) 5. If display reads “INSERT CARD,” do so. ( ) 6. Air blank completed.
( ) 7. Standard Calibration Check completed. Test results 0. AC. ( ) 8. Air blank completed.
( ) 9. When display reads “TEST COMPLETE,” remove printed record. Attach the record to the completed checklist. ( ) 10. Type “Q” and press the ENTER key on the keyboard.
SIGNATURE