EXHIBIT F-1 OPERATIONAL CHECKLIST
All data is extracted from pdf, click here to view the pdf.
-
ARIZONA DEPARTMENT OF PUBLIC SAFETY
STANDARD OPERATIONAL PROCEDURE INTOXILYZER MODEL 5000EN
DUPLICATE BREATH TEST
SUBJECT NAME DATE
AGENCY OPERATOR
INSTRUMENT SERIAL # LOCATION
TEST RESULTS
0.
AC
TIME
0.
AC
TIME
0.
AC
TIME
Immediately preceding administration of the tests, subject underwent at least a 15-minute deprivation period:
From to
by
(Time) (Time) (Name)
( ) 1. Display reads “PUSH BUTTON TO START TEST” or “PRESS START TEST BUTTON TO START NEXT TEST.” Ensure
breath tube is warm to touch. ( ) 2. Press START TEST button.
( ) 3. If display reads “INSERT CARD,” do so. ( ) 4. Input information in response to display. ( ) 5. Air blank completed.
( ) 6. If the display reads “IS SIMULATOR SOLUTION TEMPERATURE 34° C ± 0.2° C?,” check temperature using thermometer, type “Y” or “N;” verify Concurrent Calibration Check completed.
( ) 7. Insert the mouthpiece in the breath tube. Have the subject blow as long as possible. Record AC result above. ( ) 8. Air blank completed.
( ) 9. a. If display reads “WAIT,” go to Step 11
OR
b. If display reads “TEST COMPLETE,” go to Step 10
OR
c. If display reads “IS SIMULATOR SOLUTION TEMPERATURE 34° C ± 0.2° C?,” check temperature using thermometer, type “Y” or “N;” verify Concurrent Calibration Check completed, go to Step 10.
( ) 10. When the display reads “TEST COMPLETE,” remove the test record. ( ) 11. Repeat Steps 1 through 9, as necessary.
Note: Duplicate breath tests shall be administered at intervals of not less than 5 nor more than 10 minutes apart and the two consecutive tests shall agree within 0.020 alcohol concentration.