THIS REPORT PREPARED UNDER DUTY IMPOSED BY A.A.C. R13-10-104(A) ARIZONA DEPARTMENT OF PUBLIC SAFETY
STANDARD QUALITY ASSURANCE PROCEDURES INTOXILYZER MODEL 5000EN
STANDARD CALIBRATION CHECK PROCEDURE (OPTION P)
1. a. Ensure the dry gas tank is attached to the instrument and contains a standard alcohol concentration solution alcohol standard.
b. Pour a standard alcohol concentration solution into a clean dry simulator and assemble the simulator. Ensure that a tight seal is made. Turn on the simulator and allow temperature to reach 34° C ± 0.2° C.
2. Intoxilyzer 5000EN display reads “PUSH BUTTON….”
3. Ensure Intoxilyzer 5000EN calibration standard is set for “G” for gas or “W” for wetbath.
4. Type “C” and press the ENTER key on the keyboard.
6. Calibration check completed.
8. When display reads “TEST COMPLETE,” type “Q” and ENTER on the keyboard.