EXHIBIT E-4  


Latest version.

All data is extracted from pdf, click here to view the pdf.

  • THIS REPORT PREPARED PURSUANT TO DUTY IMPOSED BY A.A.C. R13-10-104(A) ARIZONA DEPARTMENT OF PUBLIC SAFETY

    STANDARD QUALITY ASSURANCE PROCEDURES INTOXILYZER MODEL 5000*

    STANDARD QUALITY ASSURANCE PROCEDURE

    QA SPECIALIST                                                                              AGENCY                                                                                                  

    DATE                                                                                                  TIME                                                                                                         

    INTOXILYZER SERIAL #                                                               LOCATION                                                                                               

    (     )         1.  Display reads “READY TO START” or “PUSH BUTTON TO START TEST.”

    DIAGNOSTIC TESTS

    (     )         1.  DVM Test check. Setting should be between .010 and .600. Mode selection switch S2 on, S1 and S3 off or keyboard menu selection “H”. Reading is                                        .

    (     )         2.  Display Test check. Mode selection switch S1 on and S2 and S3 off or keyboard menu selection “V”. (    )      3.  Printer Test check. Mode selection switch S1, S2, S3 off or keyboard menu selection “P”.

    (     )         4.  Clock time check. Mode selection switch S10 on or keyboard menu selection “E”. (    )      5.  Date check. Mode selection switch S11 on or keyboard menu selection “E”.

    OPERATIONAL TESTS

    (     )         1.  Alcohol-free subject Test result 0.                        AC ( )      2.  Error recognition logic system functioning

    Invalid test printed

    (     )         3.  Proper sample recognition system Invalid sample printed

    Deficient sample printed

    (     )         4.  Standard Calibration Check standard 0.                                 AC Results: 0.                           AC

    Instrument operating accurately and properly. YES             NO          

    COMMENTS                                                                                                                                                                                                       

    SIGNATURE

    * WITH OR WITHOUT VAPOR RECIRCULATION AND WITH OR WITHOUT KEYBOARD

    DPS Form Exh E-4 (Rev 05-1)