EXHIBIT E-5  


Latest version.

All data is extracted from pdf, click here to view the pdf.

  • THIS REPORT PREPARED PURSUANT TO DUTY IMPOSED BY A.A.C. R13-10-104(A) ARIZONA DEPARTMENT OF PUBLIC SAFETY

    STANDARD QUALITY ASSURANCE PROCEDURES INTOXILYZER MODEL 5000

    STANDARD QUALITY ASSURANCE PROCEDURE (OPTION P)

    Display reads “READY TO START” or “PUSH BUTTON TO START TEST.” DIAGNOSTIC TESTS

    1.        DVM Test check. Value is between .010 and .600. Keyboard menu selection “H”.

    2.        Display Test check. Keyboard menu selection “V”.

    3.        Clock time check. Keyboard menu selection “E”.

    4.        Date check. Keyboard menu selection “E”.

    OPERATIONAL TESTS

    1.        Alcohol-free subject Test result 0.000 AC.

    2.        Error recognition logic system functioning. Invalid test printed.

    3.        Proper sample recognition system. Invalid sample printed.

    Deficient sample printed.

    4.        Calibration standard 0.100 AC.

    Instrument operating accurately and properly. Enter “Y” or “N”. DPS Form Exh E-5 (Rev 05-1)

Historical Note

New Exhibit E-5 made by final rulemaking at 12 A.A.R. 1916, effective 9:00 a.m., May 18, 2006 (Supp. 06-2).

Department of Public Safety - Alcohol Testing