THIS REPORT PREPARED PURSUANT TO DUTY IMPOSED BY A.A.C. R13-10-104(A) ARIZONA DEPARTMENT OF PUBLIC SAFETY
STANDARD QUALITY ASSURANCE PROCEDURES INTOXILYZER MODEL 5000
STANDARD CALIBRATION CHECK PROCEDURE (OPTION P)
1. Ensure simulator is on and contains a standard alcohol concentration solution of known value, 0.100 AC, at temperature of 34° C
± 0.2° C.
2. Intoxilyzer display reads “READY TO START” or “PUSH BUTTON”.
3. Set Intoxilyzer mode selection in the ACA mode by selecting “C” on the keyboard menu.
4. Press ENTER on keyboard.
5. Air blank completed.
6. Calibration check completed.
7. Confirm Standard Calibration Check reading is in 0.090 to 0.110 range.
8. Air blank completed.
9. Test complete.
Instrument reading is within acceptable accuracy limits. Enter “Y” or “N”.
DPS Form Exh E-3 (Rev 05-01)