EXHIBIT H-4


Latest version.

All data is extracted from pdf, click here to view the pdf.

  • THIS REPORT PREPARED PURSUANT TO DUTY IMPOSED BY A.A.C. R13-10-104(A) ARIZONA DEPARTMENT OF PUBLIC SAFETY

    STANDARD QUALITY ASSURANCE PROCEDURES ALCO SENSOR RBT AZ

    CALIBRATION

    AGENCY                                                                                                                                 DATE                                                                      

    QA SPECIALIST                                                                                         LOCATION                                                                                     

    RBT AZ SERIAL #                                                                ALCO-SENSOR AZ SERIAL #                                                                           

    (     )         1. Have a standard alcohol concentration solution ready.

    This may be a simulator (at 34° C ± 0.2° C) or a dry gas alcohol standard. Standard value: 0.                              AC. (      )         2. Depress RBT AZ ON button.

    (     )         3. Depress Time button, enter PIN #, depress #1 button. ( )      4. Follow RBT AZ and AS AZ display instructions.

    (     )         5. Device temperature registers between 23° C and 27° C.

    (     )         6. After a blank reading of 0.000 is displayed and the standard value is displayed, depress F3.

    (     )         7. When AS AZ display flashes “CAL”, introduce standard for 7 seconds; depress the MANUAL button on the AS AZ at 5 seconds (while continuing to introduce the standard for another 2 seconds.)

     )         8. Remove test record when printout is complete.

    (     )         9. Run a calibration check on the Standard Calibration Check Procedure.

    Test results:                                   AC.

    COMMENTS                                                                                                                                                                                                       

    SIGNATURE

    DPS Form Exh H-4 (Rev 05-01)