THIS REPORT PREPARED PURSUANT TO DUTY IMPOSED BY A.A.C. R13-10-104(A) ARIZONA DEPARTMENT OF PUBLIC SAFETY
STANDARD QUALITY ASSURANCE PROCEDURES ALCO SENSOR RBT AZ
STANDARD QUALITY ASSURANCE PROCEDURE
AGENCY DATE
QA SPECIALIST LOCATION
RBT AZ SERIAL # ALCO-SENSOR AZ SERIAL #
( ) 1. Have a standard alcohol concentration solution ready.
This may be a simulator (at 34° C ± 0.2° C) or a dry gas alcohol standard. Standard value: 0. AC. ( ) 2. Depress RBT AZ ON button.
Depress Time button. Enter PIN #.
Depress zero button.
( ) 3. Follow RBT AZ and AS AZ display instructions.
( ) 4. Device temperature registers between 10° C and 40° C.
( ) 5. When AS AZ display reads “CHEK”, introduce standard for 7 seconds; depress the MANUAL button on the AS AZ at 5 seconds (while continuing to introduce the standard for another 2 seconds.)
( ) 6. Test results 0. AC.
( ) 7. Remove test record when printout is complete. ( ) 8. Turn off RBT AZ.
( ) 1. Date and time correct.
( ) 2. Alcohol-free subject test result 0. AC. ( ) 3. Proper sample recognition system.
( ) 4. Fuel cell response time for a standard solution.
Standard value: AC. Time sec.
( ) 5. Controls, displays, and printer worked correctly during the above quality assurance procedures.
COMMENTS
SIGNATURE