Arizona Administrative Code (Last Updated: November 17, 2016) |
![]() |
Title 13. PUBLIC SAFETY |
![]() |
Chapter 10. DEPARTMENT OF PUBLIC SAFETY ALCOHOL TESTING |
![]() |
Article 1. DETERMINATION OF ALCOHOL CONCENTRATION |
EXHIBIT H-2
All data is extracted from pdf, click here to view the pdf.
-
THIS REPORT PREPARED PURSUANT TO DUTY IMPOSED BY A.A.C. R13-10-104(A) ARIZONA DEPARTMENT OF PUBLIC SAFETY
STANDARD QUALITY ASSURANCE PROCEDURES ALCO SENSOR RBT AZ
STANDARD CALIBRATION CHECK PROCEDURE
AGENCY DATE
QA SPECIALIST LOCATION
RBT AZ SERIAL # ALCO SENSOR AZ SERIAL #
( ) 1. Have a standard alcohol concentration solution ready.
This may be a simulator (at 34° C ± 0.2° C) or a dry gas alcohol standard. Standard value: 0. AC. ( ) 2. Depress RBT AZ ON button.
Depress Time button. Enter PIN #.
Depress zero button.
( ) 3. Follow RBT AZ and AS AZ display instructions.
( ) 4. Device temperature registers between 10° C and 40° C.
( ) 5. When AS AZ display reads “CHEK”, introduce standard for 7 seconds; depress the MANUAL button on the AS AZ at 5 seconds (while continuing to introduce the standard for another 2 seconds.)
( ) 6. Test results 0. AC.
( ) 7. Remove test record when printout is complete. ( ) 8. Turn off RBT AZ.
COMMENTS
SIGNATURE