EXHIBIT H-1 OPERATIONAL CHECKLIST
All data is extracted from pdf, click here to view the pdf.
-
ARIZONA DEPARTMENT OF PUBLIC SAFETY
STANDARD OPERATIONAL PROCEDURE ALCO SENSOR RBT AZ
SUBJECT NAME DATE
AGENCY OPERATOR
LOCATION
RBT AZ SERIAL # ALCO SENSOR AZ SERIAL #
TEST RESULTS
0. AC
TIME
0. AC
TIME
0. AC
TIME
Immediately preceding administration of the tests, subject underwent at least a 15-minute deprivation period:
From to
by
(Time) (Time) (Name)
( ) 1. Depress RBT AZ ON button.
( ) 2. Depress zero set button, select subject or quick test. ( ) 3. Follow RBT AZ and AS AZ display instructions.
( ) 4. Enter case # &/or DL # if required.
( ) 5. Device temperature registers between 10° C and 40° C. ( ) 6. a. If quick test, go to step 7.
b. If subject test, repeat steps 3 – 6 for duplicate test.
c. If the second subject test is not within 0.020 of the first test, repeat steps 3-6.
d. If the second subject test is within 0.020 of the first test, go to step 7.
e. If the third subject test, go to step 7.
( ) 7. Remove test record when printout is complete. ( ) 8. Turn off RBT AZ.
Note: Duplicate breath tests shall be administered at intervals of not less than 5 nor more than 10 minutes and the two consecutive tests shall agree within 0.020 alcohol concentration.