EXHIBIT E-6 OPERATIONAL CHECKLIST
All data is extracted from pdf, click here to view the pdf.
-
ARIZONA DEPARTMENT OF PUBLIC SAFETY STANDARD OPERATIONAL PROCEDURE
INTOXILYZER MODEL 5000 - WITHOUT VAPOR RECIRCULATION
AND WITHOUT KEYBOARD DUPLICATE BREATH TEST
SUBJECT NAME DATE
AGENCY OPERATOR
INSTRUMENT SERIAL # LOCATION
TIME OF TEST
TEST RESULTS
0.
AC
TIME
0.
AC
TIME
0.
AC
TIME
Immediately preceding administration of the tests, subject underwent at least a 15-minute deprivation period:
From to
by
(Time) (Time) (Name)
( ) 1. Display reads “READY TO START” or “PUSH BUTTON TO START TEST”. Breath tube is warm to touch. ( ) 2. Press Start Test button.
( ) 3. If display reads “INSERT CARD”, do so. ( ) 4. Air blank completed.
( ) 5. Insert mouthpiece into breath tube. Have subject blow as long as possible. Record results above. ( ) 6. Air blank completed.
( ) 7. a. If display reads “WAIT”, go to step 8.
OR
b. If display reads “TEST COMPLETE”, go to step 9. ( ) 8. Repeat steps 1 through 7.
( ) 9. When display reads “TEST COMPLETE”, remove test record card. If duplicate tests have not been obtained between 5 and 10 minutes apart with a .020 AC agreement, repeat steps 1 through 7.
Note: Duplicate breath tests shall be administered at intervals of not less than 5 nor more than 10 minutes apart and the two consecutive tests shall agree within 0.020 alcohol concentration.