THIS REPORT PREPARED PURSUANT TO DUTY IMPOSED BY A.A.C. R13-10-104(A) ARIZONA DEPARTMENT OF PUBLIC SAFETY
STANDARD QUALITY ASSURANCE PROCEDURES INTOXILYZER MODEL 5000*
STANDARD CALIBRATION CHECK PROCEDURE
QA SPECIALIST AGENCY
DATE TIME
INTOXILYZER SERIAL # LOCATION
( ) 1. Pour a standard alcohol concentration solution into a clean dry simulator jar and assemble the simulator. Ensure that a tight seal has been made. Standard value: 0. AC.
( ) 2. Turn on the simulator and allow the temperature to reach 34° C ± 0.2° C.
( ) 3. Set Intoxilyzer mode selection in the ACA mode by switching mode selection switch #9 on or selecting “C” on keyboard menu.
( ) 4. Attach simulator to the simulator entrance port on the Intoxilyzer.
( ) 5. Intoxilyzer 5000 display reads “READY TO START” or “PUSH BUTTON”. ( ) 6. Push Start Test button or press enter on keyboard.
( ) 7. Insert card in response to display. ( ) 8. Air blank completed.
( ) 9. Standard Calibration Check completed. Test results 0. AC. ( ) 10. Air blank completed.
( ) 11. When test complete, remove printed record. Attach the record to the completed checklist.
( ) 12. Return mode selection switch #9 to off position after all calibration checks are complete or type Q and enter on keyboard. SIGNATURE
* WITH OR WITHOUT VAPOR RECIRCULATION AND WITH OR WITHOUT KEYBOARD