  • THIS REPORT PREPARED PURSUANT TO DUTY IMPOSED BY A.A.C. R13-10-104(A) ARIZONA DEPARTMENT OF PUBLIC SAFETY

    STANDARD QUALITY ASSURANCE PROCEDURES INTOXILYZER MODEL 5000*

    STANDARD CALIBRATION CHECK PROCEDURE

    QA SPECIALIST                                                                                          AGENCY                                                                                       

    DATE                                                                                              TIME                                                                                                              

    INTOXILYZER SERIAL #                                                           LOCATION                                                                                                    

    (     )         1.  Pour a standard alcohol concentration solution into a clean dry simulator jar and assemble the simulator. Ensure that a tight seal has been made. Standard value: 0.                                              AC.

    (     )         2.  Turn on the simulator and allow the temperature to reach 34° C ± 0.2° C.

    (     )         3.  Set Intoxilyzer mode selection in the ACA mode by switching mode selection switch #9 on or selecting “C” on keyboard menu.

    (     )         4.  Attach simulator to the simulator entrance port on the Intoxilyzer.

    (     )         5.  Intoxilyzer 5000 display reads “READY TO START” or “PUSH BUTTON”. (      )         6.  Push Start Test button or press enter on keyboard.

    (     )         7.  Insert card in response to display. (      )         8.  Air blank completed.

    (     )         9.  Standard Calibration Check completed. Test results 0.                                      AC. (      )       10.  Air blank completed.

    (     )       11.  When test complete, remove printed record. Attach the record to the completed checklist.

    (     )       12.  Return mode selection switch #9 to off position after all calibration checks are complete or type Q and enter on keyboard. SIGNATURE                                                                                                                                                                                                         