Section R12-1-743. Therapy-related Computer Systems
All data is extracted from pdf, click here to view the pdf.
-
The licensee shall perform acceptance testing on the treatment plan- ning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing shall include, as applicable, verification of:
1. The source-specific input parameters required by the dose calculation algorithm;
2. The accuracy of dose, dwell time, and treatment time cal- culations at representative points;
3. The accuracy of isodose plots and graphic displays;
4. The accuracy of the software used to determine sealed source positions from radiographic images; and
5. The accuracy of electronic transfer of the treatment deliv- ery parameters to the treatment delivery unit from the treatment planning system.