Section R12-1-733. Dosimetry Equipment  


Latest version.

All data is extracted from pdf, click here to view the pdf.

  • A.      Except for low dose-rate remote afterloader sources where the source output or activity is determined by the manufacturer, a licensee shall have a calibrated dosimetry system available for use. To satisfy this requirement, one of the following two con- ditions shall be met.

    1.        The system shall have been calibrated using a system or source traceable to the National Institute of Science and Technology (NIST) and published protocols accepted by nationally recognized bodies; or by a calibration labora- tory accredited by the American Association of Physicists in Medicine (AAPM). The calibration shall have been performed within the previous two years and after any servicing that may have affected system calibration; or

    2.        The system shall have been calibrated within the previous four years. Eighteen to 30 months after that calibration, the system shall have been intercompared with another dosimetry system that was calibrated within the past 24 months by NIST or by a calibration laboratory accredited by the AAPM. The results of the intercomparison shall indicate that the calibration factor of the licensee’s system had not changed by more than two percent. The licensee may not use the intercomparison result to change the cali- bration factor. When intercomparing dosimetry systems to be used for calibrating sealed sources for therapeutic units, the licensee shall use a comparable unit with beam attenuators or collimators, as applicable, and sources of the same radionuclide as the source used at the licensee’s facility.

    B.       The licensee shall have a dosimetry system available for use for spot-check output measurements, if applicable. To satisfy this requirement, the system may be compared with a system that has been calibrated in accordance with subsection (A). This comparison shall have been performed within the previ- ous year and after each servicing that may have affected sys- tem calibration. The spot-check system may be the same system used to meet the requirement in subsection (A).

    C.      The licensee shall retain, for three years from the date of the procedure, a record of each calibration, intercomparison, and comparison.

Historical Note

New Section made by final rulemaking at 13 A.A.R.

1217, effective May 5, 2007 (Supp. 07-1).