Section R12-1-730. Installation, Maintenance, Adjustment, and Repair of an Afterloader Unit, Teletherapy Unit, or Gamma Stereotactic Radiosurgery Unit  


Latest version.

All data is extracted from pdf, click here to view the pdf.

  • A.      Only a person specifically licensed by the Agency, NRC, or an Agreement State shall install, maintain, adjust, or repair a remote afterloader unit, teletherapy unit, or gamma stereotac- tic radiosurgery unit that involves work on any source shield- ing, the source’s driving unit, or other electronic or mechanical component that could expose a source, reduce the shielding around a source, or compromise the radiation safety of a unit or a source.

    B.       Except for low dose-rate remote afterloader units, only a per- son specifically licensed by the Agency, NRC, or an Agree- ment State shall install, replace, relocate, or remove a sealed source or source contained in other remote afterloader units, teletherapy units, or gamma stereotactic radiosurgery units.

    C.      For a low dose-rate remote afterloader unit, only a person spe- cifically licensed by the Agency, NRC, or an Agreement State or an authorized medical physicist shall install, replace, relo- cate, or remove a sealed source contained in the unit.

    D.      A licensee shall retain a record of the installation, mainte- nance, adjustment, and repair of remote afterloader units, tele- therapy units, and gamma stereotactic radiosurgery units for three years from the completion date of the activity listed in this Section.