A. Registrants are subject to the following requirements for Intra-
oral dental radiographic units designed to be operated as a hand-held unit:
1. For all uses:
a. Operators of hand-held intraoral dental radiographic units shall be specifically trained to operate such equipment.
b. A hand-held intraoral dental radiographic unit shall be held without any motion during a patient exam- ination. A tube stand may be utilized to immobilize a hand-held intraoral dental radiographic unit during patient examination.
c. The operator shall ensure there are no bystanders within a radius of at least six feet from the patient being examined with a hand-held intraoral radio- graphic unit.
2. Additional requirements for operatories in permanent facilities:
a. Hand-held intraoral dental radiographic units shall be used for patient examinations in dental operato- ries that meet the structural shielding requirements specified by the Agency or by a qualified health or medical physicist.
b. Hand-held intraoral dental radiographic units shall not be used for patient examinations in hallways and waiting rooms.
B. Hand-held units may only be used in a manner as specified on the registration issued by the Agency.